Size | SKU | Price |
---|---|---|
65L / 2.3 Cu ft | JT:AAH26016U |
$1,923
|
150L / 5.3 Cu ft | JT:AAH26026U |
$2,391
|
205L / 7.2 Cu ft | JT:AAH26036U |
$3,044
|
Affordable yet exceptionally reliable incubators featuring film type heaters.
Models: IB-01E/11E/21E
Features:
Cu ft. | 2.3, 5.3, 7.2 |
Max Temp | 60°C |
Min Temp | Ambient +3°C |
Temp Uniformity | ± 0.2°C, ± 0.3°C,± 0.1°C |
Interior Outlets | No |
Alarms | Yes |
Datalogging Capable | No |
Warranty (years) | 1 (2 years after submission of warranty registration card) |
Site | Floor, Stackable |
Weight (lbs) | 73, 104, 126 |
Model | IB-01E | IB-11E | IB-21E | |
---|---|---|---|---|
Chamber volume (L / cu ft) | 65 / 2.3 | 150 / 5.3 | 205 / 7.2 | |
Temperature | Range (℃ / ℉) | Amb. +3 to 60 / Amb. +5.4 to 140 | ||
Fluctuation 1) at 37℃ (±℃ / ℉) | 0.2 / 0.36 | 0.3 / 0.54 | 0.1 / 0.18 | |
Variation 1) at 37℃ (±℃ / ℉) | 0.6 / 1.08 | 0.8 / 1.44 | 1.1 / 1.98 | |
Dimension (W×D×H) | Interior (mm / inch) |
423×355×445
/ 16.7×14×17.5
|
500×515×585
/ 19.7×20.3×23
|
632×515×630
/ 24.9×20.3×24.8
|
Exterior (mm / inch) |
533×475×725
/ 21×18.7×28.5
|
595×630×865
/ 23.5×24.8×34
|
740×630×910
/ 29.1×24.8×35.8
|
|
No. of wire shelf (standard/ max.) | 2 / 10 | 2 / 13 | 2 / 14 | |
Net weight (Kg / lbs) | 33 / 72.8 | 47 / 103.6 | 57 / 125.7 | |
Electrical requirements (120V, 50 / 60Hz) | 2.1A | 4.2A | 4.6A |
Do the IB Economy Series Incubators have an access port?
Yes, they do.
Are the Jeio Tech IB Economy Series Incubators heated by forced air convection or natural convection?
These are natural convection incubators; there is no fan inside the chamber.
What is the difference between Fluctuation and Variation in the spec sheet?
Variation (Temp Control Accuracy): After installing the temperature sensor in accordance with DIN12880 standard (ex. 9 pieces under 50L, 27 pieces if over 50L) After the temperature in the chamber is stabilized, after measuring for 1 hour, the most It is the difference between low and high temperature. In other words, you can see the temperature uniformity of the incubator.
Fluctuation (Uniformity): After installing the temperature sensor according to DIN12880 standard (ex. 9 pcs below 50L, 27 pcs for 50L or more) After measuring for 1 hour after the temperature in the chamber is stabilized, the lowest This is the maximum value of the deviation between temperature and high temperature. In other words, you can see the temperature stability of the incubator.